مشخصات پژوهش

صفحه نخست /Ultra-trace determination of ...
عنوان
Ultra-trace determination of silver in water samples by electrothermal atomic absorption spectrometry after preconcentration with a ligand-less cloud point extraction methodology
نوع پژوهش مقاله چاپ شده
کلیدواژه‌ها
Silver; Cloud point extraction; Electrothermal atomic absorption spectrometry; PONPE 7.5; Water samples
چکیده
A very simple and ligand-less cloud point extraction (CPE) methodology for the preconcentration of ultra-trace amounts of silver as a prior step to its determination by electrothermal atomic absorption spectrometry (ETAAS) has been developed. The method is based on the extraction of silver at pH 9 by using non-ionic surfactant polyethyleneglycolmono-p-nonylphenylether (PONPE 7.5) without adding any chelating agent Several important variables that affect the CPE efficiency and ETAAS signal were investigated and optimized. The preconcentration of 15 ml sample solution allowed us to achieve an enhancement factor of 60. The calibration graph using the preconcentration system was linear in the range of 5–100 ng l−1 with a correlation coefficient of 0.9991. The lower limit of detection (3s) obtained in the optimal conditions was 1.2 ng l−1 The relative standard deviation (R.S.D.) for eight replicate determinations at 30 ng l−1 Ag level was 4.2%. The proposed method was successfully applied to the ultra-trace determination of silver in water samples.
پژوهشگران جمشید منظوری لشکر (نفر اول)، حسین عبدالمحمدزاده (نفر دوم)