مشخصات پژوهش

صفحه نخست /Performance evaluation of a ...
عنوان
Performance evaluation of a proliferation chamber with external tirred conditioning tank for expansion of a suspendable stem cell model
نوع پژوهش مقاله چاپ شده
کلیدواژه‌ها
Proliferation chamber External stirred conditioning tank Suspendable stem cell Shear stress Cell roundness
چکیده
Performance of a proliferation chamber with external stirred conditioning tank (PC-ESCT) was evaluated to expand a cell line as a model of suspendable stem cells. A hematopoietic cell line (U937) showed considerable cell proliferation in the PC-ESCT at seeding densities of 104 and 105 cells/mL (97.5 ± 1.7 and 24.5 ± 1.9-fold), as compared to the cultures in T-flask (80 ± 2.8 and 21.5 ± 1.4-fold) and stirred tank bioreactor (STB) (no detectable increase and 2.85 ± 0.8-fold) after 8 days of culture. The round-shaped cells appeared dominantly in the T-flask and PC-ESCT cultures, while few cells could maintain their original phenotype or survive in the STB system during the culture period. The cell surface marker analysis of CD14 for testing self-renewing proliferation of the cells also coincide with the cell appearance and quantitative morphological analysis based on the roundness parameter. The present study showed the potential of the PC-ESCT for the expansion of suspendable stem cells in vitro.
پژوهشگران یونس بیگی خسروشاهی (نفر اول)، علی برادر خوش فطرت (نفر دوم)، زهرا ابوالقاسمی (نفر سوم)، کریم شمس اسنجان (نفر چهارم)